Test Method |
Moving Specimens up and down by elevating system |
Temperature |
Test |
Low Range |
-55℃ ~ -20℃ |
Chamber |
High Range |
60℃ ~ 150℃ |
(Basket) |
Accuracy |
±1.0℃ at 100℃ |
|
Uniformity |
±2.0℃℃ at 100℃ |
|
Recovery time |
Within 5min |
Low |
Range |
-70℃ ~ 0℃(Pre-cooling temp.) |
Chamber |
Accuracy |
±2.0℃ |
|
Pull-down time |
RT → -70℃ within 70min |
High |
Range |
60℃ ~ 200℃(Pre-heating temp.) |
Chamber |
Accuracy |
±2.0 |
|
Heat up time |
RT → 200℃ within 20min |
Material |
Internal |
Stainless steel 1.5t |
External |
Steel (SS41, Powder coating) |
Insulation |
Mineral wool 100t |
Heater |
Incoloy Pipe heater |